Digital Stainless Steel N95 Mask Testing Equipment, Model Name/Number: Nre-co2 M-01, 1% Of Fs
₹10,00,000
Seller : New Ray Engineering
Specification
- size
- 1 X 0.6 X 1.2 MTR
- weight
- 300 KGS
- accuracy
- 1% OF FS
- capacity
- 10% OF CO2
- material
- Stainless Steel
- display type
- Digital
- physical state
- Solid
- model name number
- NRE-CO2 M-01
Add Quantity
Description
The� following� screen� appear once� the� system� is switched� on a. co2 content� test:- once you select �the� above� test �following� screen apears 1. Set� the� parameters� as� per the� �instructions below note ; the� approximate� setting� are� as� follows, initially the� trial� is� taken without� mask to� normalise� the� machine ( 1. Co2 flow :� 1. 8- 2lpm, 2. Humidifier� bath temp: 50-52 deg c ,� 3. Cycles/ min (stoke): 25 ) 2. Once the paramet� ers are� set , you� can press the� test start" button 3. The� machine will�� start without� �co2 flow� for� a minute and� then the� co2 flow �will start� as per the setting 4. Co2 flow� is so� adjusted to� achieve� exhalation� co2�� is about 4. 85 -5. 1%�� ans� inhalation co2� is� about� 0. 4-0. 5% , keeping the� duct� in front of the hood� with blower on. 5. Once the exhalation air� temp reaches about� 37 deg� c , the mask can be� put� on the� hood ,� seal it from all� sides� and take� trials� for� exhalation (i. E. About >5 %) and� inhalation co2 %������ (i. E. > 1%) the� test� can be� repeated for� other batch of products b. simulated wearing �test when you select� this� test� the� following screen areas 1. Set the parameter� for wearing mask test 2. Run the� test� with� test start button 3. Bath temp to be� adjusted� to achieve exhalationtemp� to 37deg�� c.